Easy Dial Oxygen Regulator — Class II medical device recall Z-0272-2016
Terminated recall by Precision Medical, Inc., reported 2015-11-25, distributed in AL, AZ, CA, GA, ID, IN, KY, LA…. These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arr
| Recall number | Z-0272-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Precision Medical, Inc., Northampton, PA, United States |
| Recall initiated | 2015-10-08 |
| Classified | 2015-11-13 |
| Reported by FDA | 2015-11-25 |
| Terminated | 2016-10-27 |
| Distributed | AL, AZ, CA, GA, ID, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NY, OH, OK, PA, TX |
| Quantity | 315 |
| Reason classes | labeling |
| Lot numbers | 072215, 072315 |
Product
Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.
Reason for recall
These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0272-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.