Data Foundry

Medical device recalls 2015

Easy Dial Oxygen Regulator — Class II medical device recall Z-0272-2016

Terminated recall by Precision Medical, Inc., reported 2015-11-25, distributed in AL, AZ, CA, GA, ID, IN, KY, LA…. These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arr

Recall numberZ-0272-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPrecision Medical, Inc., Northampton, PA, United States
Recall initiated2015-10-08
Classified2015-11-13
Reported by FDA2015-11-25
Terminated2016-10-27
DistributedAL, AZ, CA, GA, ID, IN, KY, LA, MA, MI, MN, MO, MT, NC, ND, NY, OH, OK, PA, TX
Quantity315
Reason classeslabeling
Lot numbers072215, 072315

Product

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Reason for recall

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).