VITROS Chemistry Products PHYT Slides — Class II medical device recall Z-0273-2018

Terminated recall by Ortho-Clinical Diagnostics, reported 2018-01-03, distributed in PR. The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product

Recall numberZ-0273-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2017-06-12
Classified2017-12-22
Reported by FDA2018-01-03
Terminated2020-08-11
DistributedPR
Countries outside the USAU, BM, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, PL, RU, SG, VE
Reason classespotency
UPC / GTIN / UDI-DI10758750004690
Model numbers829, 8671

Product

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0273-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.