Data Foundry

Medical device recalls 2012

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System… — Class II medical device recall Z-0278-2013

Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2012-11-21, distributed nationwide. Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperat

Recall numberZ-0278-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmTerumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States
Recall initiated2012-10-16
Classified2012-11-13
Reported by FDA2012-11-21
Terminated2013-07-11
DistributedNationwide (US)
Quantity19 units
Serial numbers1041, 1083, 1084, 1101-1116
Model numbers809810

Product

HX2 Temp Manage SYS 120V The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Reason for recall

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0278-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.