Acetabular Pressurizer — Class II medical device recall Z-0280-2014
Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-11-20, distributed nationwide. On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522
| Recall number | Z-0280-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation, Portage, MI, United States |
| Recall initiated | 2013-11-01 |
| Classified | 2013-11-14 |
| Reported by FDA | 2013-11-20 |
| Terminated | 2014-02-10 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA, CH, PL |
| Reason classes | packaging |
| Lot numbers | 08183012, 08310012, 09237012, 10243012, 11109012, 11353012, 12216012 |
| Model numbers | 0206-522-000, 08183012, 08310012, 09237012, 10243012, 11109012, 11353012, 12216012 |
Product
Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is an accessory to the Advanced Cement Mixer (ACM). The Acetabular Pressurizer Tiplet P/N 0206-522-012 is sealed in a two pouches, sterilized, and labeled as part number 0206-522-000. This Acetabular Pressurizer is an accessory used to pressurize the cement applied to the prosthesis to be implanted during the surgery. The Acetabular Pressurizer is an injection molded component made with resin material DOW HDPD 10462N which is a rigid plastic once it is molded.
Reason for recall
On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manufacturing operator noticed that some assemblies were cracked. As a result of the injection pressure of the cement gun, a cracked tiplet may release from the Hub during cement applica
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0280-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.