Acetabular Pressurizer — Class II medical device recall Z-0280-2014

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2013-11-20, distributed nationwide. On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522

Recall numberZ-0280-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2013-11-01
Classified2013-11-14
Reported by FDA2013-11-20
Terminated2014-02-10
DistributedNationwide (US)
Countries outside the USCA, CH, PL
Reason classespackaging
Lot numbers08183012, 08310012, 09237012, 10243012, 11109012, 11353012, 12216012
Model numbers0206-522-000, 08183012, 08310012, 09237012, 10243012, 11109012, 11353012, 12216012

Product

Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is an accessory to the Advanced Cement Mixer (ACM). The Acetabular Pressurizer Tiplet P/N 0206-522-012 is sealed in a two pouches, sterilized, and labeled as part number 0206-522-000. This Acetabular Pressurizer is an accessory used to pressurize the cement applied to the prosthesis to be implanted during the surgery. The Acetabular Pressurizer is an injection molded component made with resin material DOW HDPD 10462N which is a rigid plastic once it is molded.

Reason for recall

On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manufacturing operator noticed that some assemblies were cracked. As a result of the injection pressure of the cement gun, a cracked tiplet may release from the Hub during cement applica