Data Foundry

Medical device recalls 2019

Liquid Protein Calibrators — Class II medical device recall Z-0283-2020

Terminated recall by Randox Laboratories Ltd., reported 2019-11-13, distributed nationwide. The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British st

Recall numberZ-0283-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRandox Laboratories Ltd., Crumlin (North), N/A, Ireland
Recall initiated2019-08-09
Classified2019-11-04
Reported by FDA2019-11-13
Terminated2021-01-08
DistributedNationwide (US)
Quantity100 kits
UPC / GTIN / UDI-DI05055273204032
Lot numbers1748, 1752IT, 1813, 1817IT, 440973, 451414, 470531, 480859
Model numbersIT2691

Product

Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems

Reason for recall

The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0283-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.