Liquid Protein Calibrators — Class II medical device recall Z-0283-2020
Terminated recall by Randox Laboratories Ltd., reported 2019-11-13, distributed nationwide. The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British st
| Recall number | Z-0283-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Randox Laboratories Ltd., Crumlin (North), N/A, Ireland |
| Recall initiated | 2019-08-09 |
| Classified | 2019-11-04 |
| Reported by FDA | 2019-11-13 |
| Terminated | 2021-01-08 |
| Distributed | Nationwide (US) |
| Quantity | 100 kits |
| UPC / GTIN / UDI-DI | 05055273204032 |
| Lot numbers | 1748, 1752IT, 1813, 1817IT, 440973, 451414, 470531, 480859 |
| Model numbers | IT2691 |
Product
Liquid Protein Calibrators (SP CAL (LIQ)) Catalogue Number: IT2691 GTIN: 05055273204032 Product Usage: Liquid Protein Calibrators are intended for in vitro diagnostic use in the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgA, IgG, IgM, Prealbumin, and Transferrin assays on Clinical Chemistry and Immunoassay systems, Abbott Spectrum, Abbott Aeroset, Abbott Architect i2000, Architect i2000sr, Ace analyser, Bayer Advia 1650, Advia 2400, Advia 1200, Dade Dimension RXL, Dimension AR, Hitachi 704, Hitachi 717, Hitachi 911, Hitachi 917, Hitachi 912, Hitachi 747, Kone progress, AU800, AU600, AU400, AU2700, AU5400, Selectra Vitalab, Synchron CX4, Synchron CX5, Synchron CX7, Synchron LX20, ILAB300, ILAB900, ILAB1800, ILAB600, RX Daytona, RX Imola, Cobas Mira, Cobas Mira S, Cobas Mira Plus systems
Reason for recall
The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.