OmniaSecure Model 3930M — Class II medical device recall Z-0285-2027
Ongoing recall by Medtronic, Inc., reported 2026-11-04, distributed nationwide. Affected leads require additional engineering evaluation due to complaints regarding difficulty in achieving f
| Recall number | Z-0285-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic, Inc., Mounds View, MN, United States |
| Recall initiated | 2026-09-17 |
| Classified | 2026-10-23 |
| Reported by FDA | 2026-11-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, CA, DE, ES, FI, FR, GB, GR, IE, IL, IT, JP, NZ, PL, SE, SI |
| Quantity | 717 |
| UPC / GTIN / UDI-DI | 00199150001170, 00199150001187, 00763000643430, 00763000643447, 00763000643454, 00763000643461, 00763000759919 |
| Serial numbers | 3930M79, CDD401008V, CDD401100V, CDD401499V, CDD401655V, CDD401951V, CDD402630V, CDD402652V, CDD402658V, CDD402681V, CDD403969V, CDD403996V, CDD404002V, CDD404005V, CDD404010V, CDD404013V, CDD404049V, CDD404050V, CDD404068V, CDD404095V, CDD404096V, CDD404102V, CDD404103V, CDD404104V, CDD404107V and 475 more |
| Model numbers | 3930M |
Product
OmniaSecure Model 3930M, LEAD 3930M74 74 CM DF4 MR 3930M74 LEAD 3930M79 79 CM DF4 MR 3930M79 The Medtronic OmniaSecure" Model 3930M lead is a steroid eluting, integrated bipolar, nonretractable screw-in, catheter delivered, transvenous ventricular lead with an RV defibrillation coil electrode. The lead is designed for pacing, sensing, cardioversion, and defibrillation therapies.
Reason for recall
Affected leads require additional engineering evaluation due to complaints regarding difficulty in achieving fixation during implantation. Potential harm of tissue trauma due to implant difficulty.
Machine access