Data Foundry

Medical device recalls 2012

ABX PENTRA Creatinine 120 CP — Class III medical device recall Z-0289-2013

Terminated recall by Horiba Instruments, Inc dba Horiba Medical, reported 2012-11-21, distributed nationwide. This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REA

Recall numberZ-0289-2013
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHoriba Instruments, Inc dba Horiba Medical, Irvine, CA, United States
Recall initiated2012-09-14
Classified2012-11-14
Reported by FDA2012-11-21
Terminated2012-11-29
DistributedNationwide (US)
Quantity8,503 units
Reason classespotency
Lot numbers203273706, 203273707, 203273709, 203273808
Model numbersA11A01868

Product

ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709. The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives. ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This reagent is intended for use on the ABX PENTRA 400 Clinical Chemistry Analyzer.

Reason for recall

This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0289-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.