Data Foundry

Medical device recalls 2016

RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod — Class II medical device recall Z-0290-2017

Terminated recall by RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.), reported 2016-10-26, distributed nationwide. RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight

Recall numberZ-0290-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.), Marquette, MI, United States
Recall initiated2016-09-16
Classified2016-10-19
Reported by FDA2016-10-26
Terminated2017-06-01
DistributedNationwide (US)
Quantity66
Reason classespotency, packaging
UPC / GTIN / UDI-DI00846468027179
Lot numbers234829, 246678
Model numbers23-CC-ROD-240

Product

RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Usage: When intended to promote fusion of the cervical spine and occipitocervico-thoracic junction (occiput-T3), the Streamline OCT Occipito-Cervico-Thoracic System is intended for: degenerative disc disease (as defined by neck or back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/ dislocation, atlanto/axial fracture with instability, occipito-cervical dislocation, deformities or curvature, tumors, pseudoarthrosis, and revision of previous cervical and upper thoracic spine surgery. The occipital bone screws are limited to occipital fixation only. The use of the pedicle screws (standard and high angle) is limited to placement in the upper thoracic spine (T1-T3) in treating thoracic conditions only. The pedicle screws are not intended for use in the cervical spine. The hooks, connectors, and rods are also intended to provide stabilization to promote fusion following reduction of fracture/ dislocation or trauma in the cervical/ upper thoracic (C1-T3) spine.

Reason for recall

RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants could cause reduced strength of these affected implants, and lead to patient injury.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0290-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.