Data Foundry

Medical device recalls 2024

Trilogy Evo Universal Ventilator — Class I medical device recall Z-0292-2025

Ongoing recall by Philips Respironics, Inc., reported 2024-11-20, distributed nationwide. Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation o

Recall numberZ-0292-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPhilips Respironics, Inc., Murrysville, PA, United States
Recall initiated2024-10-07
Classified2024-11-13
Reported by FDA2024-11-20
DistributedNationwide (US)
Quantity9,999 units
UPC / GTIN / UDI-DI00606959052000
Model numbersDS2000X11B

Product

Trilogy Evo Universal Ventilator. Model Number DS2000X11B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.

Reason for recall

Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.