Millenium MG/MC — Class II medical device recall Z-0297-2026
Ongoing recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, reported 2025-10-29. GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Servi
| Recall number | Z-0297-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, Haifa, Israel |
| Recall initiated | 2025-09-12 |
| Classified | 2025-10-23 |
| Reported by FDA | 2025-10-29 |
| Quantity | 1,386 units |
Product
Millenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Reason for recall
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0297-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.