Data Foundry

Medical device recalls 2018

AXIOM Multix MT — Class II medical device recall Z-0298-2019

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2018-11-07, distributed in AL, AR, AZ, CA, CO, DC, FL, GA…. In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, whic

Recall numberZ-0298-2019
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2018-08-06
Classified2018-10-30
Reported by FDA2018-11-07
Terminated2020-04-10
DistributedAL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, WY
Quantity11
Serial numbers1074, 1119, 1121, 1132, 1136, 1137, 1138, 1472, 1525, 1529, 1531
Model numbers8395381, 8395399, 8395415

Product

AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.

Reason for recall

In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0298-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.