AXIOM Multix MT — Class II medical device recall Z-0298-2019
Terminated recall by Siemens Medical Solutions USA, Inc, reported 2018-11-07, distributed in AL, AR, AZ, CA, CO, DC, FL, GA…. In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, whic
| Recall number | Z-0298-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Medical Solutions USA, Inc, Malvern, PA, United States |
| Recall initiated | 2018-08-06 |
| Classified | 2018-10-30 |
| Reported by FDA | 2018-11-07 |
| Terminated | 2020-04-10 |
| Distributed | AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NC, ND, NE, NM, NV, NY, OH, OK, PA, PR, SC, SD, TX, UT, VA, WI, WY |
| Quantity | 11 |
| Serial numbers | 1074, 1119, 1121, 1132, 1136, 1137, 1138, 1472, 1525, 1529, 1531 |
| Model numbers | 8395381, 8395399, 8395415 |
Product
AXIOM Multix MT (model number 8395415 or 8395399 or 8395381), a patient table used in conjunction with the Siemens Multix PRO- and TOP Systems radiographic exposures of the entire body.
Reason for recall
In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.