Data Foundry

Medical device recalls 2025

Millenium MPS — Class II medical device recall Z-0300-2026

Ongoing recall by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, reported 2025-10-29. GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Servi

Recall numberZ-0300-2026
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmGE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, Haifa, Israel
Recall initiated2025-09-12
Classified2025-10-23
Reported by FDA2025-10-29
Quantity54 units

Product

Millenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Reason for recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0300-2026. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.