Data Foundry

Medical device recalls 2026

Philips Azurion S — Class II medical device recall Z-0300-2027

Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2026-11-04, distributed nationwide. Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various

Recall numberZ-0300-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmPHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands
Recall initiated2026-10-06
Classified2026-10-27
Reported by FDA2026-11-04
DistributedNationwide (US)
Countries outside the USAE, AF, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, KY, LB, LK, LT, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VN, ZA
Quantity1 units
UPC / GTIN / UDI-DI00884838125681
Model numbers722285

Product

Philips Azurion S. Model Number: 722285.

Reason for recall

Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0300-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-10. Not affiliated with or endorsed by FDA. Not medical or legal advice.