Data Foundry

Medical device recalls 2026

EMAG system Model/Catalog Number: 418591/418591U Product Description:… — Class II medical device recall Z-0301-2027

Ongoing recall by BioMerieux SA, reported 2026-11-04, distributed nationwide. Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly

Recall numberZ-0301-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBioMerieux SA, Marcy L'Etoile, France
Recall initiated2026-09-11
Classified2026-10-27
Reported by FDA2026-11-04
DistributedNationwide (US)
Countries outside the USAT, BE, CA, DE, DK, FR, GB, NL, NO, SE
Quantity33
Reason classessoftware
Serial numbersIM03034, IM03071, IM03092, IM03104, IM03131, IM03148, IM03153, IM03175, IM03177, IM03193, IM03194, IM03240, IM03275, IM03283, IM03314, IM03321, IM03329, IM03343, IM03350, IM03351, IM03376, IM03378, IM03380, IM03414, IM03439 and 8 more
Model numbers418591/418591U

Product

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

Reason for recall

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2