Mobile Detector Holder — Class II medical device recall Z-0311-2018
Terminated recall by Philips Electronics North America Corporation, reported 2018-01-10. If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secure
| Recall number | Z-0311-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Electronics North America Corporation, Andover, MA, United States |
| Recall initiated | 2017-07-14 |
| Classified | 2018-01-03 |
| Reported by FDA | 2018-01-10 |
| Terminated | 2020-06-05 |
| Quantity | 412 |
Product
Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.
Reason for recall
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.