Mobile Detector Holder — Class II medical device recall Z-0311-2018

Terminated recall by Philips Electronics North America Corporation, reported 2018-01-10. If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secure

Recall numberZ-0311-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2017-07-14
Classified2018-01-03
Reported by FDA2018-01-10
Terminated2020-06-05
Quantity412

Product

Mobile Detector Holder - Stationary X-ray System (RF System). suitable for positioning the wireless portable detector free in the room for X-ray exposures. It supports the wireless portable detector with or without a grid.

Reason for recall

If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.