Data Foundry

Medical device recalls 2018

Medfusion — Class II medical device recall Z-0314-2018

Terminated recall by Smiths Medical ASD Inc., reported 2018-01-10, distributed nationwide. Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect

Recall numberZ-0314-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSmiths Medical ASD Inc., Minneapolis, MN, United States
Recall initiated2017-08-22
Classified2018-01-04
Reported by FDA2018-01-10
Terminated2025-04-14
DistributedNationwide (US)
Countries outside the USCA, DE, GB, IN, OM, PA, PH
Quantity9,787 units
NDC03500-0600-51
Model numbers3500

Product

Medfusion, Syringe Infusion Pump, Software Version 6 (V6), Rx Only, Manufactured by Smiths Medical ASD, Inc., 1500 Eureka Park, Lower Pemberton, Ashford, Kent, TN25 4BF, UK. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids.

Reason for recall

Potential for the syringe barrel clamp mechanism to loosen, decreasing the pump's ability to accurately detect the diameter of the syringe barrel, resulting in the unit to initiate the "Invalid Syringe Size" alarm.

Machine access