Precision Pointer Electromagnetic — Class II medical device recall Z-0314-2027
Ongoing recall by Stryker Leibinger GmbH & Co. KG, reported 2026-11-04, distributed nationwide. Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may
| Recall number | Z-0314-2027 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Stryker Leibinger GmbH & Co. KG, Freiburg Im Breisgau, Germany |
| Recall initiated | 2026-09-24 |
| Classified | 2026-10-29 |
| Reported by FDA | 2026-11-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | AT, CA, CH, DE, DK, ES, FI, FR, GB, IE, IT, JP, KR, NO, PT |
| Quantity | 513 |
| Reason classes | device malfunction |
| UPC / GTIN / UDI-DI | 07613327413205 |
| Lot numbers | 7000413361 |
Product
Precision Pointer Electromagnetic (8000-050-001)
Reason for recall
Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0314-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-10. Not affiliated with or endorsed by FDA. Not medical or legal advice.