Data Foundry

Medical device recalls 2026

Carevo — Class II medical device recall Z-0315-2027

Ongoing recall by ARJOHUNTLEIGH POLSKA Sp. z.o.o., reported 2026-11-04, distributed nationwide. The side support of the shower trolley may open unintentionally when the handles are not fully locked and addi

Recall numberZ-0315-2027
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmARJOHUNTLEIGH POLSKA Sp. z.o.o., Komorniki, Poland
Recall initiated2026-09-25
Classified2026-10-29
Reported by FDA2026-11-04
DistributedNationwide (US)
Quantity8,780 units
Reason classesdevice malfunction
Serial numbersP0347696, P0347697, P0347698, P0348019, P0348020, P0351695, P0351696, P0351697, P0351698, P0351699, P0351700, P0351701, P0351702, P0351703, P0351704, P0359775, P0359776, P0359777, P0364832, P0366288, P0368638, P0370606, P0370607, P0370608, P0370609 and 473 more
Model numbersBAC1111-01

Product

Carevo, Height-adjustable shower trolley, Model Number BAC1111-01

Reason for recall

The side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient handling. If the side support is not fully locked and an additional force is applied by the caregiver on the handle (e.g., during turning or repositioning), the panel may open unintentionally, which can lead to a resident/patient fall and serious injury, such as head injury, fractures, or other trauma resulting from a patient fall.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0315-2027. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-11-10. Not affiliated with or endorsed by FDA. Not medical or legal advice.