Data Foundry

Medical device recalls 2015

Philips HeartStart MRx Monitor/Defibrillator — Class II medical device recall Z-0320-2016

Terminated recall by Philips Electronics North America Corporation, reported 2015-12-02, distributed nationwide. The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earli

Recall numberZ-0320-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2015-10-14
Classified2015-11-20
Reported by FDA2015-12-02
Terminated2018-09-13
DistributedNationwide (US)
Countries outside the USAF, AR, AT, AU, AW, BD, BE, BG, BH, BO, BR, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, LB, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SV, TH, TN, TR, TT, TW, UA, VE, VN
Quantity81,161
Reason classessoftware

Product

Philips HeartStart MRx Monitor/Defibrillator; M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6. For use for the termination of ventricular tachycardia and ventricular fibrillation.

Reason for recall

The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an abnormal state when the device is turned off.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0320-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.