Philips HeartStart MRx Monitor/Defibrillator — Class II medical device recall Z-0320-2016
Terminated recall by Philips Electronics North America Corporation, reported 2015-12-02, distributed nationwide. The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earli
| Recall number | Z-0320-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Electronics North America Corporation, Andover, MA, United States |
| Recall initiated | 2015-10-14 |
| Classified | 2015-11-20 |
| Reported by FDA | 2015-12-02 |
| Terminated | 2018-09-13 |
| Distributed | Nationwide (US) |
| Countries outside the US | AF, AR, AT, AU, AW, BD, BE, BG, BH, BO, BR, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, LB, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SV, TH, TN, TR, TT, TW, UA, VE, VN |
| Quantity | 81,161 |
| Reason classes | software |
Product
Philips HeartStart MRx Monitor/Defibrillator; M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6. For use for the termination of ventricular tachycardia and ventricular fibrillation.
Reason for recall
The following MRx software issue has been identified: MRx model M3535A with software version F.03.06 and earlier, and model M3536A with version T.00.05 and earlier may stop the automated Ready-For-Use (RFU) test in an abnormal state when the device is turned off.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0320-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.