Data Foundry

Medical device recalls 2024

Hillrom Welch Allyn XScribe Cardiac Stress Testing System — Class II medical device recall Z-0325-2025

Ongoing recall by Baxter Healthcare Corporation, reported 2024-11-13. There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system recei

Recall numberZ-0325-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBaxter Healthcare Corporation, Deerfield, IL, United States
Recall initiated2024-10-11
Classified2024-11-07
Reported by FDA2024-11-13
Quantity766 units
UPC / GTIN / UDI-DI00113480185300, 00116070311448, 00116240458430, 00116430533831, 00116470539367, 00116490541265, 00116490541753, 00116500543944, 00117080554665, 00117160563860, 00117270577498, 00117390588114, 00117410589558, 00117470596305, 00118120623358, 00118120623365, 00118140625486, 00118140625493, 00118140625738, 00118160628177, 00118180635070, 00118230000780, 00118260000798, 00118270000498, 00118300000047 and 122 more
Serial numbers111130075526, 113030149726, 113030149727, 113100154916, 113240166924, 113480185300, 114150206492, 114160206692, 114340225721, 114450235208, 115170254512, 115170254513, 115340269178, 115350269606, 116070308454, 116070311445, 116070311446, 116070311447, 116070311448, 116110347144, 116110347146, 116110348774, 116110348790, 116120365527, 116120365528 and 475 more

Product

Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer

Reason for recall

There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0325-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.