Data Foundry

Medical device recalls 2023

Colonoscope: for endoscopy and endoscopic surgery within the lower… — Class II medical device recall Z-0329-2024

Ongoing recall by Aizu Olympus Co., Ltd., reported 2023-11-29, distributed nationwide. Channel air drying process was not validated, and that a small percentage of scopes returned to customers afte

Recall numberZ-0329-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmAizu Olympus Co., Ltd., Aizuwakamatsu, N/A, Japan
Recall initiated2023-09-25
Classified2023-11-17
Reported by FDA2023-11-29
DistributedNationwide (US)
Quantity3,344 units
UPC / GTIN / UDI-DI04953170305115, 04953170305139, 04953170305153, 04953170305177, 04953170307621, 04953170307638, 04953170307645, 04953170307652, 04953170307669, 04953170307690, 04953170307706, 04953170312397, 04953170312410, 04953170334160, 04953170334184, 04953170339585, 04953170339608, 04953170339639, 04953170339653
Model numbers510, CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL

Product

Colonoscope: for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve). Model Numbers: CF-2T160I, CF-2T160L, CF-H170I, CF-H170L, CF-H180AI, CF-H180AL, CF-H180DI, CF-H180DL, CF-H190I, CF-H190L, CF-HQ190I, CF-HQ190L, CF-LV1I, CF-LV1L, CF-Q150I, CF-Q150L, CF-Q160S, CF-Q180AI, CF-Q180AL. 510(k) Numbers: K954451, K112680, K100584, K051645, K111756, K001241, K911278.

Reason for recall

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection