Vascutek Gelsoft Vascular Graft — Class II medical device recall Z-0336-2014
Terminated recall by Terumo Cardiovascular Systems Corporation, reported 2013-11-27, distributed in CA, NC. Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graf
| Recall number | Z-0336-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Terumo Cardiovascular Systems Corporation, Ann Arbor, MI, United States |
| Recall initiated | 2013-10-23 |
| Classified | 2013-11-18 |
| Reported by FDA | 2013-11-27 |
| Terminated | 2014-02-27 |
| Distributed | CA, NC |
| Quantity | 9 |
| Reason classes | labeling |
| Serial numbers | 0001765391, 0001765392, 0001765393, 0001765394, 0001765395, 0001765396, 0001765397, 0001765398, 0001765399 |
| Model numbers | 631508 |
Product
Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT***STERILE***EO***Item No: 631508***Diameter 8mm***Usable Length 15cm***Manufactured By***VASCUTEK Ltd. a TERUMO Company Newmains Avenue Inchinnan, Renfrewshire PA4 9RR Scotland***" The Vascutek Gelsoft Vascular Prosthesis is a gelatin-sealed, woven polyester graft indicated for abdominal and peripheral vascular repair, i.e. replacement or bypass in aneurysmal and occlusive disease of arteries
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0336-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.