Cardiosave Hybrid IntraAortic Balloon Pump — Class I medical device recall Z-0336-2022

Ongoing recall by Datascope Corp., reported 2021-12-22, distributed nationwide. Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short v

Recall numberZ-0336-2022
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDatascope Corp., Mahwah, NJ, United States
Recall initiated2021-11-15
Classified2021-12-15
Reported by FDA2021-12-22
DistributedNationwide (US)
UPC / GTIN / UDI-DI10607567108391, 10607567108414, 10607567108421, 10607567108438, 10607567109008, 10607567109053, 10607567109107, 10607567111117, 10607567111940, 10607567113432, 10607567114187
Model numbers0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-32, 0998-UC-0800-33, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-0800-36, 0998-UC-0800-45, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55, 0998-UC-0800-65, D998

Product

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Reason for recall

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.