Cardiosave Hybrid IntraAortic Balloon Pump — Class I medical device recall Z-0336-2022
Ongoing recall by Datascope Corp., reported 2021-12-22, distributed nationwide. Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short v
| Recall number | Z-0336-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Datascope Corp., Mahwah, NJ, United States |
| Recall initiated | 2021-11-15 |
| Classified | 2021-12-15 |
| Reported by FDA | 2021-12-22 |
| Distributed | Nationwide (US) |
| UPC / GTIN / UDI-DI | 10607567108391, 10607567108414, 10607567108421, 10607567108438, 10607567109008, 10607567109053, 10607567109107, 10607567111117, 10607567111940, 10607567113432, 10607567114187 |
| Model numbers | 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-32, 0998-UC-0800-33, 0998-UC-0800-34, 0998-UC-0800-35, 0998-UC-0800-36, 0998-UC-0800-45, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55, 0998-UC-0800-65, D998 |
Product
Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.
Reason for recall
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0336-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.