Data Foundry

Medical device recalls 2018

ECG Out Cable — Class II medical device recall Z-0348-2018

Terminated recall by Philips Electronics North America Corporation, reported 2018-01-24. Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary b

Recall numberZ-0348-2018
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPhilips Electronics North America Corporation, Andover, MA, United States
Recall initiated2017-07-17
Classified2018-01-14
Reported by FDA2018-01-24
Terminated2018-07-12
Countries outside the USCA
Quantity15,811
Model numbers989803195641, M1783A, M5526A

Product

ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 October 8, 2014

Reason for recall

Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs): " On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave. " On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*. *Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.