Data Foundry

Medical device recalls 2022

Trinity Biotech HbA1c — Class III medical device recall Z-0350-2023

Terminated recall by Primus Corporation dba Trinity Biotech, reported 2022-12-07. The Control Level I material value was showing results exceeding the upper limit of the acceptable range after

Recall numberZ-0350-2023
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPrimus Corporation dba Trinity Biotech, Kansas City, MO, United States
Recall initiated2021-10-05
Classified2022-12-01
Reported by FDA2022-12-07
Terminated2023-08-09
Countries outside the USAE, AT, CL, IN, IT, KR, PE, PK, RO, TH, TR, VN
Quantity182 kits
Lot numbers12161
Model numbers01-04-0020, 05-01-0040B, A1C

Product

Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis

Reason for recall

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0350-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.