Data Foundry

Medical device recalls 2019

LAG SCREW 3 — Class II medical device recall Z-0351-2020

Terminated recall by Smith & Nephew, Inc., reported 2019-11-20, distributed nationwide. LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide

Recall numberZ-0351-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSmith & Nephew, Inc., Memphis, TN, United States
Recall initiated2019-10-07
Classified2019-11-08
Reported by FDA2019-11-20
Terminated2023-03-08
DistributedNationwide (US)
Countries outside the USAE, BE, BR, CN, ES, FR, GB, IN, JP, RU, SG, ZA
Quantity116 devices
Lot numbers18DM11613A, 18HM02575, 18HM13327A, 18HM20956A
Model numbers71674032

Product

LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.

Reason for recall

LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.