Data Foundry

Medical device recalls 2024

The IMAGEnet 6 Ophthalmic Data System is a software program that is… — Class II medical device recall Z-0362-2025

Ongoing recall by Topcon Medical Systems, Inc., reported 2024-11-20, distributed nationwide. The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "

Recall numberZ-0362-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmTopcon Medical Systems, Inc., Oakland, NJ, United States
Recall initiated2024-09-17
Classified2024-11-12
Reported by FDA2024-11-20
DistributedNationwide (US)
Quantity3,144 units
UPC / GTIN / UDI-DI04562123288582
Serial numbersAB9070101, AB9070102, AB9070103, AB9070104, AB9070105, AB9070107, AB9070108, AB9070109, AB9070110, AB9070111, AB9070112, AB9070113, AB9070114, AB9070115, AB9070117, AB9070118, AB9070119, AB9070121, AB9070122, AB9070123, AB9070124, AB9070126, AB9070127, AB9070129, AB9070130 and 475 more

Product

The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices. It is intended for processing and displaying ophthalmic images and optical coherence tomography data. The IMAGEnet 6 Ophthalmic Data System uses the same algorithms and reference databases from the original data capture device as a quantitative tool for the comparison of posterior ocular measurements to a database of known normal subjects.

Reason for recall

The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0362-2025. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.