Data Foundry

Medical device recalls 2013

Siemens Chemistry Calibrator — Class II medical device recall Z-0365-2014

Terminated recall by Siemens Healthcare Diagnostics, reported 2013-11-27, distributed nationwide. Siemens Healthcare Diagnostics is conducting a field correction for Siemens Chemistry Calibrator Lots 090445B

Recall numberZ-0365-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Healthcare Diagnostics, Tarrytown, NY, United States
Recall initiated2013-09-24
Classified2013-11-19
Reported by FDA2013-11-27
Terminated2018-05-17
DistributedNationwide (US)
Countries outside the USAE, AR, AT, BE, BG, CA, CH, CL, CN, CO, CZ, DE, DK, EE, EG, ES, FR, GB, GR, GT, HR, ID, IL, IT, KR, LT, LU, LV, MX, NL, NO, PE, PL, PT, RS, SE, SG, SI, SK, TH, TN, TR, TW, UY, VN
Reason classespotency
Lot numbers090445B
Model numbers09784096, T03-1291-62
Expiration dates2014-07, 2014-10

Product

Siemens Chemistry Calibrator (CAL) used with the ADVIA Chemistry Direct Bilirubin_2 Reagent, Siemens Material Number (SMN) 10312279, REF 09784096, Part Number T03-1291-62; IVD Product usage: Usage: DBIL_2: For in vitro diagnostic use in the quantitative determination of direct bilirubin in human serum and plasma on the ADVIA Chemistry systems. Such measurements are useful in the screening of liver function disorders or in the diagnosis of jaundice.

Reason for recall

Siemens Healthcare Diagnostics is conducting a field correction for Siemens Chemistry Calibrator Lots 090445B and 090445C on the ADVIA 1200, 1650, 1800, and 2400 Chemistry Systems. The System Specific Values (SSVs) for the Direct Bilirubin Assay (DBIL_2) have been reassigned for these lots. Siemens has determined that the DBIL_2 SSVs currently assigned to these lots may cause an average shift of

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0365-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.