Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil — Class II medical device recall Z-0365-2025

Ongoing recall by Stryker Corporation, reported 2024-11-20, distributed nationwide. There is a potential for the product to activate without manual input when plugged into a power source or rema

Recall numberZ-0365-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmStryker Corporation, Portage, MI, United States
Recall initiated2024-10-04
Classified2024-11-13
Reported by FDA2024-11-20
DistributedNationwide (US)
Countries outside the USAU, CA, JP
Quantity1,120,860 eaches
UPC / GTIN / UDI-DI07613327117202
Lot numbers53863666, 53863667, 53863668, 53863669, 53863670, 53863671, 54403959, 54403960, 54403961, 54403962, 54403963, 54403964, 54403965, 54409851, 54409852, 54409853, 54409854, 54409855, 54409856, 54409857, 54409858, 54409859, 54409860, 54409862, 54409919, 54409920, 54409921, 54409922, 54409923, 54409924, 54409925, 54409926, 54409927, 54533880, 54533881, 54533882, 54533883, 54533884, 54654973, 54665296 and 143 more
Model numbers0703-047-000

Product

Stryker Neptune SafeAir Rocker-Switch Smoke Evacuation Pencil, Coated, 70mm - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-047-000

Reason for recall

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries