Data Foundry

Medical device recalls 2016

Zimmer¿ Air Dermatome II Handpiece Loaner — Class II medical device recall Z-0368-2017

Terminated recall by Zimmer Surgical Inc, reported 2016-11-16, distributed nationwide. Zimmer Biomet Surgical has received product complaints reporting that the surface coating applied to the devic

Recall numberZ-0368-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer Surgical Inc, Dover, OH, United States
Recall initiated2016-07-04
Classified2016-11-04
Reported by FDA2016-11-16
Terminated2017-06-21
DistributedNationwide (US)
Quantity1,030 units
Lot numbers01-8851-001-00
Model numbers00-8851-001-00, 01-8851-001-00

Product

Zimmer¿ Air Dermatome II Handpiece Loaner, item number 01-8851-001-00. Device is packaged inside of a padded-foam corrugate box. A product label is applied to the lid of the box. The width plates, when sold individually, are placed in a labeled poly zipper bag, which is then placed into a corrugated shipper carton or padded envelope.

Reason for recall

Zimmer Biomet Surgical has received product complaints reporting that the surface coating applied to the device is blistering, peeling and discoloring after usage over time (see Image 1). There have been no injuries associated with the reported complaints. In addition, an independent assessment determined that the blistering, peeling and potentially flaking of the surface coating is not expected to result in any long range health consequences. Given that the frequency of this surface coating condition can worsen with age, Zimmer Biomet Surgical is currently making preparations for replacement activities to follow in order to prevent future complaints. This document is provided to alert all users of the potential for the surface coating to exhibit blistering, peeling, and discoloring after usage over time. A separate recall removal notice will be issued to facilities, in phases, with detailed instructions beginning in August 2016. Your sales representatives will be notified when a replacement is ready. Do not return product at this time as part of this action. Image 1.