Data Foundry

Medical device recalls 2021

Stealthstation System w/ Stealthstation Cranial Software 3 — Class I medical device recall Z-0383-2022

Ongoing recall by Medtronic Navigation, Inc., reported 2021-12-29, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized wit

Recall numberZ-0383-2022
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic Navigation, Inc., Louisville, CO, United States
Recall initiated2021-11-11
Classified2021-12-22
Reported by FDA2021-12-29
DistributedAK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Countries outside the USGB
Quantity2,587
Reason classessoftware
Model numbers9733763, 9735585

Product

Stealthstation System w/ Stealthstation Cranial Software 3.1.1,3.1.2,3.1. or SynergyCranial Software. Synergy Cranial, Model: 9733763, and StealthStation Cranial, Model: 9735585

Reason for recall

Cranial biopsy procedure software can enter a state where the biopsy depth gauge is no longer synchronized with the rest of the navigational information on the screen and may display an incorrect position of the biopsy needle, which may result in prolonged procedure, the need for additional surgical procedure, tissue injury.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0383-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.