Data Foundry

Medical device recalls 2019

ORTHOFIX Catalogue Number: ref 1-1355001 — Class II medical device recall Z-0385-2020

Terminated recall by Orthofix Srl, reported 2019-11-20, distributed in AZ, CA, FL, GA, HI, IA, IL, IN…. As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail

Recall numberZ-0385-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrthofix Srl, Veronaverona, N/A, Italy
Recall initiated2018-09-27
Classified2019-11-14
Reported by FDA2019-11-20
Terminated2022-02-01
DistributedAZ, CA, FL, GA, HI, IA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV
Countries outside the USAE, AU, BE, BR, CH, CL, DE, DK, ES, FR, GB, IE, IL, IT, JP, KW, LK, LT, NL, PE, PL, SA, SG, ZA
Lot numbers273348, 289816, 300243, 300244
Model numbers1-1355001

Product

ORTHOFIX Catalogue Number: ref 1-1355001, DRILL BIT D.2,7 MM L.127 MM TIN COATED- QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18054242S10918

Reason for recall

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0385-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.