Data Foundry

Medical device recalls 2019

ORTHOFIX Catalogue Number: ref: 99-92503 — Class II medical device recall Z-0392-2020

Terminated recall by Orthofix Srl, reported 2019-11-20, distributed in AZ, CA, FL, GA, HI, IA, IL, IN…. As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail

Recall numberZ-0392-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrthofix Srl, Veronaverona, N/A, Italy
Recall initiated2018-09-27
Classified2019-11-14
Reported by FDA2019-11-20
Terminated2022-02-01
DistributedAZ, CA, FL, GA, HI, IA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV
Countries outside the USAE, AU, BE, BR, CH, CL, DE, DK, ES, FR, GB, IE, IL, IT, JP, KW, LK, LT, NL, PE, PL, SA, SG, ZA
UPC / GTIN / UDI-DI18033509855962
Lot numbersB1207287
Model numbers99-92503

Product

ORTHOFIX Catalogue Number: ref: 99-92503, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE LONG SCREWS, RX ONLY, UDI: (01) 18033509855962

Reason for recall

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.