Data Foundry

Medical device recalls 2012

00780302601 MIS 12/14 Slotted Femoral Head Provisional -3 — Class II medical device recall Z-0396-2013

Terminated recall by Zimmer, Inc., reported 2012-11-28, distributed nationwide. Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inf

Recall numberZ-0396-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmZimmer, Inc., Warsaw, IN, United States
Recall initiated2010-10-10
Classified2012-11-19
Reported by FDA2012-11-28
Terminated2014-04-30
DistributedNationwide (US)
Countries outside the USBB, CA, MX
Quantity56,516 all
Lot numbers69998600

Product

00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS 12/14 Slotted Femoral Head Provisional +3.5X26 00780302604 MIS 12/14 Slotted Femoral Head Provisional +7X26 00780302605 MIS 12/14 Slotted Femoral Head Provisional +10.5X26 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Reason for recall

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat