Data Foundry

Medical device recalls 2021

SYNCHRON Systems Cholesterol — Class II medical device recall Z-0399-2022

Ongoing recall by Beckman Coulter Inc., reported 2021-12-29, distributed nationwide. Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low

Recall numberZ-0399-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2021-11-10
Classified2021-12-20
Reported by FDA2021-12-29
DistributedNationwide (US)
Countries outside the USCA, IN, LB, MX, PH, SA, SG, TH, ZA
Quantity1,433
Reason classesspecification failure, device malfunction
UPC / GTIN / UDI-DI15099590575113
Lot numbersM106450
Model numbers467825

Product

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Reason for recall

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0399-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.