Smith & Nephew Taylor Spatial Frame Website software — Class II medical device recall Z-0409-2018
Terminated recall by Smith & Nephew, Inc., reported 2018-01-31. In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the f
| Recall number | Z-0409-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Smith & Nephew, Inc., Memphis, TN, United States |
| Recall initiated | 2017-12-13 |
| Classified | 2018-01-24 |
| Reported by FDA | 2018-01-31 |
| Terminated | 2020-10-05 |
| Reason classes | software |
| Model numbers | 71070401 |
Product
Smith & Nephew Taylor Spatial Frame Website software, Catalog Number 71070401. A web-based and mobile application software component of a multilateral external fixation system that is intended for fracture fixation, fixation of long bones and for joint fusions and limb lengthening or deformity corrections which involve cutting of the bone.
Reason for recall
In rare circumstances, when surgeons use the software to plan the strut adjustment, an inverted image of the frame is generated by the software. The frame will appear inverted in the preview images on the Strut Settings tab.