KING LAD SILI CONE SINGLE USE — Class II medical device recall Z-0419-2013
Terminated recall by King Systems Corp., reported 2012-11-28, distributed nationwide. The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization in
| Recall number | Z-0419-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | King Systems Corp., Noblesville, IN, United States |
| Recall initiated | 2012-11-02 |
| Classified | 2012-11-21 |
| Reported by FDA | 2012-11-28 |
| Terminated | 2013-05-28 |
| Distributed | Nationwide (US) |
| Lot numbers | K120618 |
| Model numbers | LAD-303 |
Product
KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD¿ Patient Size: 30-50kg Product Usage: Laryngeal airway
Reason for recall
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO