Data Foundry

Medical device recalls 2022

da Vinci SP Surgical systems with the following Model Name/ Model… — Class II medical device recall Z-0419-2022

Ongoing recall by Intuitive Surgical, Inc., reported 2022-01-05, distributed nationwide. Issue was identified during internal engineering evaluation. Use of the system with the affected software vers

Recall numberZ-0419-2022
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmIntuitive Surgical, Inc., Sunnyvale, CA, United States
Recall initiated2021-09-22
Classified2021-12-27
Reported by FDA2022-01-05
DistributedNationwide (US)
Quantity46
Reason classessoftware
UPC / GTIN / UDI-DI00886874114605, 00886874114735, 00886874114742

Product

da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A

Reason for recall

Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.