da Vinci SP Surgical systems with the following Model Name/ Model… — Class II medical device recall Z-0419-2022
Ongoing recall by Intuitive Surgical, Inc., reported 2022-01-05, distributed nationwide. Issue was identified during internal engineering evaluation. Use of the system with the affected software vers
| Recall number | Z-0419-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Intuitive Surgical, Inc., Sunnyvale, CA, United States |
| Recall initiated | 2021-09-22 |
| Classified | 2021-12-27 |
| Reported by FDA | 2022-01-05 |
| Distributed | Nationwide (US) |
| Quantity | 46 |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00886874114605, 00886874114735, 00886874114742 |
Product
da Vinci SP Surgical systems with the following Model Name/ Model Number / UDI: ASSY,PSS,SP1098,P4/ 380601-44/ 00886874114605 ASSY,SSC,SP1098,P4/ 380940-44 / 00886874114735 ASSY,VSS,SP1098,P4/ 380941-44 / 00886874114742 SOFTWARE, SYSTEM EMBEDDED RLS, SP1098, B70_P4_B327 (OS4 v4.0.0 (P4_B327) / 610112-327 / N/A
Reason for recall
Issue was identified during internal engineering evaluation. Use of the system with the affected software version may experience either an inability to deliver energy or an inadvertent delivery of energy. Use may require surgeon to resolve an injury or need moderate intraoperative intervention. Also, use may potentially cause user frustration or a minor delay to troubleshoot.