RAPIDPOINT 500 Blood Gas System — Class II medical device recall Z-0423-2024
Ongoing recall by Siemens Healthcare Diagnostics Inc, reported 2023-12-06, distributed nationwide. There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause fa
| Recall number | Z-0423-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics Inc, Norwood, MA, United States |
| Recall initiated | 2023-09-26 |
| Classified | 2023-11-30 |
| Reported by FDA | 2023-12-06 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AR, AT, AU, BD, BE, BG, BH, BO, BR, CA, CH, CL, CN, CO, CR, CW, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JO, JP, KE, KR, KW, LB, LT, LU, LV, MX, MY, NL, NO, NP, NZ, OM, PE, PH, PK, PL, PT, PY, QA, RO, RS, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, UY, VN, ZA |
| Quantity | 12,363 units |
| UPC / GTIN / UDI-DI | 00630414589169, 00630414590844, 00630414590851, 00630414590868 |
Product
RAPIDPOINT 500 Blood Gas System
Reason for recall
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.