Data Foundry

Medical device recalls 2013

20mm addEASE — Class II medical device recall Z-0424-2014

Terminated recall by B. Braun Medical, Inc., reported 2013-12-04, distributed nationwide. Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC130

Recall numberZ-0424-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmB. Braun Medical, Inc., Allentown, PA, United States
Recall initiated2013-04-04
Classified2013-11-27
Reported by FDA2013-12-04
Terminated2014-08-26
DistributedNationwide (US)
Quantity1,746,079 units
Reason classespackaging
Lot numbers0061275133, 0061275135, 0061275136, 0061279143, 0061279144, 0061279145, 0061279146, 0061289059, 0061289060, 0061289061, 0061289062, 0061289063, 0061289064, 0061289065, 0061289066, 0061289067, 0061299664

Product

20mm addEASE(TM) Binary Connector with 17 Ga. Needle for use with PAB(R) Containers, BC2000 Product is intended for fluid transfer between a Partial Additive Bag (PAB(R)) container and a drug vial.

Reason for recall

Possible solution leakage at the interface of the addEASE(TM) Binary Connector (Product Codes BC2000 and BC1300) and the PAB(R) bag. Based on extensive root cause investigation, it has been determined that the leakage is the result of a misaligned cannula (needle) on the addEASE(TM) Binary Connector, which pierces the PAB(R) bag stopper to access the solution. The misaligned cannula may result i