Data Foundry

Medical device recalls 2012

IMRIS ORT100 — Class II medical device recall Z-0428-2013

Terminated recall by Innovative Magnetic Resonance Imaging Systems Inc., reported 2012-11-28, distributed nationwide. IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the

Recall numberZ-0428-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInnovative Magnetic Resonance Imaging Systems Inc., Winnipeg, Manitoba, Canada
Recall initiated2012-10-19
Classified2012-11-21
Reported by FDA2012-11-28
Terminated2012-12-11
DistributedNationwide (US)
Quantity23
Model numbers110470-000, 112216-00

Product

IMRIS ORT100 (Part Number 110470-000) and ORT200 (Part Number 112216-00), are tables used in both the Neuro II-SE and Neuro III-SV Magnetic Resonance Imaging Systems.

Reason for recall

IMRIS Inc. initiated a recall of their OR Tables ORT100 and ORT200. When proper procedure is not followed, the foot end of the IMRIS intra-operative OR table may be attracted to the MRI magnet as the magnet approaches the table. This poses a potential risk of injury to both the patient and operators standing near the table.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0428-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.