4 — Class II medical device recall Z-0432-2013
Terminated recall by OrthoPediatrics Corp, reported 2012-11-28, distributed nationwide. Complaint received from sales representative that the drill guide tubes would not screw into replenishment ins
| Recall number | Z-0432-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | OrthoPediatrics Corp, Warsaw, IN, United States |
| Recall initiated | 2012-10-18 |
| Classified | 2012-11-22 |
| Reported by FDA | 2012-11-28 |
| Terminated | 2013-09-18 |
| Distributed | Nationwide (US) |
| Quantity | 24 |
| Lot numbers | 7861001, 7861101 |
| Model numbers | 01-1050-0066, 01-1050-0067 |
Product
4.5mm PLEO Osteotomy Guide. OrthoPediatrics Corp. markets a pediatric, extension osteotomy bone plate system. This system is marketed under the brand name OrthoPediatrics PediLock Extension Osteotomy Plate ( PLEO ). The product included in this recall is a Class I, 4.5mm PLEO Osteotomy Guide Instrument, common name: Guide, Surgical, Instrument. The PediLoc¿ Extension Osteotomy Plate is part of the OrthoPediatrics~PediLoc¿ system. The PediLoc¿ Extension Osteotomy Plate consists of 3.5mm and 4.5mm plates, left and right side specific. The PediLoc¿ Extension Osteotomy Plate is designed to fit the anatomy of the distal femur in children and adolescents, above the distal femoral physis. The distal section of the plate is designed so that three locking screws can be placed in the distal bone segment at an angle parallel to the distal femoral physis. The proximal portion of the plate is designed to accept locking or non-locking screws. The subject instrument of this recall is an adjunct instrument for the convenience of some surgeons who choose to use it. This instrument is not illustrated in the Surgical Technique, nor required to perform the procedure.
Reason for recall
Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.