Data Foundry

Medical device recalls 2013

Assembly — Class II medical device recall Z-0436-2014

Terminated recall by Intuitive Surgical, Inc., reported 2013-12-11, distributed nationwide. Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by

Recall numberZ-0436-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmIntuitive Surgical, Inc., Sunnyvale, CA, United States
Recall initiated2013-11-11
Classified2013-12-03
Reported by FDA2013-12-11
Terminated2014-02-18
DistributedNationwide (US)
Countries outside the USAT, AU, BE, BG, BR, CA, CH, CN, DE, DK, ES, FR, GB, IL, IN, IT, JP, KR, LU, MX, NL, NO, RO, SE, SG, TH, TR, TW
Quantity1,386
Model numbers380900-02, 380900-03, IS2000/3000

Product

Assembly, Patient Side Manipulator on da Vinci Surgical Systems: da Vinci S Surgical System; da Vinci SI Surgical System da Vinci Si-e Surgical System; Manufactured by Intuitive Surgical Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS2000/3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electro cautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electro cautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system is indicated for adult and pediatric use

Reason for recall

Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall momentarily and then suddenly catch-up to the master position if the surgeon pushes through the resistance.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0436-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.