Data Foundry

Medical device recalls 2015

Optional PS500 Power Supply Unit for the Evita V500 Ventilator and… — Class I medical device recall Z-0436-2016

Terminated recall by Draeger Medical, Inc., reported 2015-12-30, distributed nationwide. The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with th

Recall numberZ-0436-2016
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDraeger Medical, Inc., Telford, PA, United States
Recall initiated2015-12-01
Classified2015-12-23
Reported by FDA2015-12-30
Terminated2020-04-03
DistributedNationwide (US)
Quantity2,422
Reason classessoftware
Model numbers8416400/8417400

Product

Optional PS500 Power Supply Unit for the Evita V500 Ventilator and Babylog VN500 Ventilator. Babylog VN500 is a ventilation unit intended for the ventilation of neonatal and pediatric patients.

Reason for recall

The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the charging algorithm in the current software. Devices used for patient transport will be a priority.