Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No… — Class II medical device recall Z-0440-2014

Terminated recall by Boston Scientific Corporation, reported 2013-12-11, distributed nationwide. The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip

Recall numberZ-0440-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBoston Scientific Corporation, Marlborough, MA, United States
Recall initiated2013-11-05
Classified2013-12-04
Reported by FDA2013-12-11
Terminated2014-12-08
DistributedNationwide (US)
Countries outside the USAU, CA, CH, DE, DK, FR, GB, IE, IT, NL, ZA
Quantity274
Lot numbers15708431, 15708432, 15708433, 15812875, 15836355, 15855545, 15876213, 15896950, 15935564, 16028474, 16044445, 16046280
Model numbers250-227

Product

Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No M0062502270, RDF Catalog No 250-227, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract.

Reason for recall

The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0440-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.