Karl Storz Udel Sterilization Tray Instructions for Use — Class II medical device recall Z-0442-2022
Ongoing recall by Karl Storz Endoscopy, reported 2022-01-05, distributed in AK, AL, AR, AZ, CA, CO, CT, DC…. To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) u
| Recall number | Z-0442-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Karl Storz Endoscopy, El Segundo, CA, United States |
| Recall initiated | 2021-10-15 |
| Classified | 2021-12-29 |
| Reported by FDA | 2022-01-05 |
| Distributed | AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY |
| Countries outside the US | BB, BM, BR, BS, BZ, CA, CW, DE, DO, EC, JP, KR, KY, MX, PA, PH, TT, VE |
| Quantity | 71,892 |
| Reason classes | chemical contamination |
Product
Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS); 39301AS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 4MM X 18CM); 39301BCTS/CAMERA TRAY (INLINE/C-MOUNT CAMERA HEADS); 39301BS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 5MM X 35CM); 39301C1S/RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE & LT CABLE); 39301CS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPE 10MM X 31CM); 39301DS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39301HCTS/CAMERA TRAY (STERRAD, STANDARD CAMERA HEADS); 39301PHTS/CAMERA TRAY (STERRAD, PENDULUM/INLINE CAM HEADS); 39305C1S/BASKET-STYLE RIGID TELESCOPE TRAY (UDEL, 1 RIGID SCOPE 4MM X 30CM); 39305C2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39305L1S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE); 39305L2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39311AS/RIGID TELESCOPE TRAY (STERRAD, 4 RIGID SCOPES & LT CABLE); 39311BS/RIGID TELESCOPE TRAY (STERRAD, TWO LAP SCOPES/CABLE); 39314FS/RIGID TELESCOPE TRAY (STERRAD, FOUR LAP SCOPES/CABLE); 39401AS/FLEXIBLE ENDOSCOPE TRAY (UP TO 39CM); 39402AS/FLEXIBLE ENDOSCOPE TRAY (FIBERSCOPES); 39403AS/FLEXIBLE ENDOSCOPE TRAY (CCD VIDESCOPES); 39406AS/FLEXIBLE ENDOSCOPE TRAY (CMOS VIDEOSCOPES)
Reason for recall
To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0442-2022. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.