Data Foundry

Medical device recalls 2020

Computed Tomography X-Ray System — Class II medical device recall Z-0446-2021

Ongoing recall by Shanghai United Imaging Healthcare Co., Ltd., reported 2020-11-18, distributed nationwide. A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP

Recall numberZ-0446-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmShanghai United Imaging Healthcare Co., Ltd., Shanghai, China
Recall initiated2020-08-05
Classified2020-11-12
Reported by FDA2020-11-18
DistributedNationwide (US)
Quantity5
UPC / GTIN / UDI-DI06971576831036
Serial numbers353020, 353021, 353023, 353045, 353049
Model numbers550

Product

Computed Tomography X-Ray System, Model uCT 550, Rx, CE, GTIN: 06971576831036 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Reason for recall

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0446-2021. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.