Anesthesia circuit kit — Class I medical device recall Z-0449-2026

Ongoing recall by Medline Industries, LP, reported 2025-11-19, distributed nationwide. Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cra

Recall numberZ-0449-2026
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedline Industries, LP, Northfield, IL, United States
Recall initiated2025-10-10
Classified2025-11-13
Reported by FDA2025-11-19
DistributedNationwide (US)
Countries outside the USMX, PA
Quantity96,406 units
Reason classespackaging
UPC / GTIN / UDI-DI10884389688588, 40884389688589
Lot numbers25EBV570, 25FBT843

Product

Anesthesia circuit kit; CIRCUIT,ANES,EXTENDFLEX,120,LF DYNJAA10230

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.