Data Foundry

Medical device recalls 2012

Neptune¿ 2 Rover — Class II medical device recall Z-0454-2013

Terminated recall by Stryker Instruments Div. of Stryker Corporation, reported 2012-12-12, distributed nationwide. Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor

Recall numberZ-0454-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmStryker Instruments Div. of Stryker Corporation, Portage, MI, United States
Recall initiated2011-05-24
Classified2012-12-03
Reported by FDA2012-12-12
Terminated2014-05-28
DistributedNationwide (US)
Countries outside the USCA, JP
Quantity5,865
Lot numbers0733700353
Model numbers0702-001-000

Product

Neptune¿ 2 Rover - ULTRA ~120V ~ 60 Hz 12A REF 0702-001-000, Rx, Caution, consult accompanying documents Consult instructions for use, Max Vacuum; 21.0 in-Hg, 530 mm-Hg, The Neptune Waste Management System consists of a mobile rover unit used to suction and collect fluid waste, small debris and electrocautery smoke from a surgical site. Mobility allows the rover unit to be relocated to a waste disposal area where the rovers collection canisters can be emptied, via the docking station. Intended to be used in the Operating Room, Surgical Centers and Doctors Offices to collect and dispose of surgical fluid waste as well as collect smoke from electrocautery or laser devices.

Reason for recall

Beginning May 24, 2011 Stryker Neptune 2 (N2) Waste Management Systems castors were replaced with a new castor design because investigation found the old design had the potential for the casters to bind compromising the Rover's ability to roll. This can result in additional strain on the operator, due to multiple exposures resulting in back pain.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0454-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.