Alaris EtCO2 Module — Class II medical device recall Z-0455-2016
Terminated recall by CareFusion 303, Inc., reported 2015-12-30. CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service usin
| Recall number | Z-0455-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | CareFusion 303, Inc., San Diego, CA, United States |
| Recall initiated | 2015-11-12 |
| Classified | 2015-12-21 |
| Reported by FDA | 2015-12-30 |
| Terminated | 2017-01-23 |
| Quantity | 157 units |
| Model numbers | 8300 |
Product
Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).
Reason for recall
CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0455-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.