Data Foundry

Medical device recalls 2015

Alaris EtCO2 Module — Class II medical device recall Z-0455-2016

Terminated recall by CareFusion 303, Inc., reported 2015-12-30. CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service usin

Recall numberZ-0455-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCareFusion 303, Inc., San Diego, CA, United States
Recall initiated2015-11-12
Classified2015-12-21
Reported by FDA2015-12-30
Terminated2017-01-23
Quantity157 units
Model numbers8300

Product

Alaris EtCO2 Module, Model 8300. A capnograph that continuously monitors end tidal carbon dioxide (EtCO2), fractional inspired carbon dioxide (FiCO2), and respiratory rate (RR).

Reason for recall

CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an incorrect measurement of CO2.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-0455-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.